Start a New Run

Start a new run
It is recommended to begin each new run by rebooting
(shutting down and restarting) the instrument, then clicking
Assist (wrench icon), then Restart Data Collection
Workflow, to bring the software to the beginning of the
workflow.
Hover over the Data Collection button
the DSP Control Center and select New/Continue Run.
If a Data Collection run is in progress, clicking New/Continue Run will take you to the current step in
the run.
Load the slides into the slide holder
1. Remove clean slide holder from the instrument and open the slide slot clamps. Inspect the
holder to ensure there is no residue from a previous run. Manually move the spring clamp to
ensure that it moves freely and is not blocked by debris. Use a lint-free wipe and dH
the slide holder, if needed. See
guidelines.
IMPORTANT:
under the gasket of the slide clamp. Avoid using slides with colored or blank labels, which may
not be detected by the imaging system. Add a dark mark to the slide label area using a
permanent lab pen to improve slide detection if needed.
2. Load slides (through step 5) one at a time to minimize exposure to air. Inspect your slide for
debris and residue.
Clean the bottom of your slide with a lint-free wipe and 70% ethanol.
l
Inspect your slide to ensure there is no residual material from the hydrophobic pen if used
l
during slide preparation.
FOR RESEARCH USE ONLY. Not for use in diagnostic procedures.
22
(see Figure 8)
Cleaning the Slide Holder on page 105
Slide labels must not be positioned beyond the frosted portion of the slide or
MAN-10152-01 for software v2.5