GeoMx DSP Instrument User Manual
GeoMx DSP Instrument Maintenance
i. Insert slide holder with a spare slide in Slot 1, covered with Buffer S.
ii. Insert the spare collection plate.
iii. Select any Morphology Marker kit from the dropdown menu. Do not select any Panel
Kit. The system will warn that you will not be able to collect from this slide.
iv. Proceed through the Preparing Instrument and Initializing Fluidics steps to generate a low
magnification scan.
b. Define a scan area in the low magnification scan to proceed to the high magnification scan.
c. When the high magnification scan is complete, the system will initiate cleaning and move to
the Collection Complete idlesoftware state.
d. Once at the Collection Complete idle software state, click on Remove Slides and
Microplate .
e. When prompted, open the door to retrieve the slide holder. Store the slide, thoroughly clean
the slide holder and return the slide holder to inside the instrument.
f. You may remove the spare collection plate (and save for similar uses).
g. When finished, close the door to the instrument.
FOR RESEARCH USE ONLY. Not for use in diagnostic procedures.
108
MAN-10152-01 for software v2.5