GeoMx DSP Instrument Maintenance

Cleaning the Slide Holder

Clean the slide holder after each run to ensure proper instrument function.
1. If neccessary, remove buffer from loaded slides with a pipette and then remove slides from the
slide holder.
2. Rinse the slide holder thoroughly with distilled water from a squirt bottle. Gently pat dry with
Kimwipes.
3. Inspect the entire slide holder. Ensure that the springs of each slide clamp and the gasket
(blue rubber) of each slide slot are clear of debris from salt, dust, or lint.
4. If neccessary, gaskets may be removed for thorough cleaning with distilled water. Allow to air
dry or pat dry with Kimwipes. Inspects for tears or damage. Carefully replace gaskets to slide
holder, ensuring a good fit.
To minimize wear and tear on gaskets, avoid frequent removal from the slide holder. A brief cleaning
following each instrument run should keep the gaskets clean without needing to remove them from
the slide holder.
5. Visually inspect the digital micromirror device (DMD) calibration target
ensure it is clear of all debris. Do not contact the target unless necessary to gently wipe away
debris. DO NOT apply pressure to the target.
6. Return the slide holder to the GeoMx instrument between runs.
IMPORTANT:
If the slide holder has been dropped, its performance could be compromised.
Run a test slide before using the slide holder with experimental samples. Contact
GeoMxSupport@nanostring.com
FOR RESEARCH USE ONLY. Not for use in diagnostic procedures.
GeoMx DSP Instrument User Manual
Figure 100: Underside of slide holder with DMD calibration
for more specific guidance.
(see Figure
(DMD cal) target indicated
100) to