nanoString GeoMx DSP User Manual page 11

Instrument
Table of Contents

Advertisement

MAN-10152-01 for software v2.5
Instrument Introduction
The GeoMx DSP instrument requires the following reagent bottles in the reagent bay
Figure
4):
Buffer S: 1 L bottles in reagent bay slots 1 and 4.
l
Buffer H: 250 mL bottles in reagent bay slots 2 and 3.
l
In addition, the waste bottle, with adequate space to collect waste, must be present in the
position to the right of the reagent bottles.
IMPORTANT:
Ensure the instrument is in an idle software state before changing reagents (see
Bring GeoMx DSP Software to an Idle Software State on page
off any reagents. Always replace entire bottles when reagents are low, using the wizard.
For information on replacing reagent bottles, see
100.
FOR RESEARCH USE ONLY. Not for use in diagnostic procedures.
Figure 4: GeoMx DSP instrument reagent bay diagram
GeoMx DSP Instrument Reagents on page
GeoMx DSP Instrument User Manual