Administration Menu
F
M
ILE
ANAGEMENT
1. Select File Management to set archive
settings
(see Figure
Remote server path: Enter the network path in UNC format for the remote location where
l
scan data will be uploaded. The remote server path must contain four backslashes, the IP
address and folder path, in this format:
\\\\server IP address\\folder path
Since archiving from the GeoMx DSP System can occur at any time, the remote location must be
available at all times.
Username: enter the username for a user with read and write privileges for the remote
l
server path entered. If including the domain, use this format: user@domain.
Password: enter the correct password for the username. Passwords may not include $ or !
l
or / .
2. Check the Archive Data box at the bottom of the form.
Do not attempt to turn on Archive Settings while a run is in progress.
3. Click the Save button. A confirmation or error message will appear.
4. After first turning on archiving, the GeoMx system will create a file titled mount-
verification.txt in the archive location. Check the archive destination folder to confirm this file
was created successfully. This indicates that archiving is working properly. If the expected
file is not present, contact
A list of dates and location of system backups can be viewed by clicking the System Backups
button under Administration.
FOR RESEARCH USE ONLY. Not for use in diagnostic procedures.
68).
GeoMxSupport@nanostring.com
GeoMx DSP Instrument User Manual
Figure 68: File Management tab under Administration
to troubleshoot.