GeoMx DSP Instrument User Manual
Start a New Run
Set up the GeoMx DSP instrument
The prompts may appear in a different order than that listed here.
Replace Plate?
This screen only appears if the system detects that a collection plate is already loaded on the
instrument.
Select Yes to replace the collection plate. Select No to use the collection plate currently loaded
on the GeoMx DSP instrument. Select Next.
Completely used collection plate: if the selected plate is already full, you will be prompted to
l
use a different plate.
Partially used collection plate: if the selected plate is only partially full, you may choose to
l
use the remaining wells for the present run. Collection will be set to use the next available well
or row on the plate (see details in
Open/Close the Door
Open the instrument door when prompted. Once the door is
open
(see Figure 12)
step.
If the door seems to be locked when it should be unlocked, push down lightly on the door, then begin to
pull it open. It should then unlatch and open.
When prompted to close the door, first ensure that the plate cover is pushed all the way to the
right. Once the door is closed, the system will confirm that the plate cover is out of the way and
lock the door.
FOR RESEARCH USE ONLY. Not for use in diagnostic procedures.
24
Identify Plate on page
, the system will display the next
MAN-10152-01 for software v2.5
25).
Figure 12: Open the GeoMx instrument