Ibp - EDAN elite V5 User Manual

Patient monitor
Table of Contents

Advertisement

Patient Monitor User Manual
Resolution
1
☆ Accuracy
Refresh Time
Temperature Calibration
Measuring Mode
Transient Response Time
Note 1: The accuracy consists of two parts, as following:
Accuracy (not including sensor): ± 0.1 ° C
Sensor accuracy: ≤ ± 0.2 ° C
A.16 PR
PR (SpO
) EDAN
2
Nellcor
PR
EDAN
(NIBP)
Omron
SunTech 30 bpm to 220 bpm
PR (IBP)
EDAN
A.17 IBP
Complies with IEC 60601-2-34: 2011.
Technique
Channel
0.1 ° C (0.1 ºF)
± 0.3 ° C
Every 1 s to 2 s
At an interval of 5 to 10 minutes
Direct Mode
≤ 30 s
Measuring range
25 bpm to 300 bpm
20 bpm to 300 bpm
40 bpm to 240 bpm
Adult/ Pediatric mode: 40
bpm to 200 bpm
Neonatal mode: 40 bpm
to 240 bpm
20 bpm to 300 bpm
Accuracy
± 2 bpm
± 3 bpm (20 bpm to 250
bpm)
±3 bpm or 3.5%, whichever
is greater
± 2 bpm or 2% of the
readings
±3 bpm or ±2%, whichever
is greater
30 bpm to 300 bpm: ± 2 bpm
± 2%,
or
greater;
20
bpm
undefined
Direct invasive measurement
8
- 258 -
Product Specifications
Resolution
1 bpm
1 bpm
1 bpm
1 bpm
1 bpm
1 bpm
whichever
is
to
29
bpm: