Patient Monitor User Manual
Message
ECG
Signal
Exceeded
ECG Noise
ECG Comm Fail
RESP
RESP Comm Fail
RESP Noise
RR Exceed
Cause
ECG measuring signal is
beyond measuring range.
ECG measuring signal is
greatly interrupted.
ECG module failure or
communication failure
RESP module failure or
communication failure
RR cannot be measured due
to patient movement.
RR measuring value is out
of the measure range.
Alarm Level
Low
Low
High
High
Low
Medium
- 55 -
Alarm Information
Action Taken
Check
lead
connection
and
patient condition
Stop
measuring
function of ECG
module, and notify
biomedical engineer
or
manufacturer's
service staff.
Stop
measuring
function of RESP
module, and notify
biomedical engineer
or
the
manufacturer's
service staff.
Check whether the
RESP leads are well
connected. Keep the
patient
calm
for
better monitoring.
Check
whether
interference to the
respiratory
signal
exists. And check
whether the patient
is
breathing
normally; breathing
too rapidly or too
slowly
may
endanger
patient's
life.