Patient Monitor User Manual
Measure Parameters
Unit
☆ Measuring Range
☆ EtCO
2
☆ FiCO
2
☆ AwRR
Resolution
☆ EtCO
Accuracy
2
☆ AwRR Accuracy
Operation Mode
Sample
Gas
(sidestream)
O
Compensation
2
Range
Resolution
Default
Barometric
Compensation
Anesthetic Gas Compensation
Range
Resolution
Default
EtCO
, FiCO
2
mmHg, %, Kpa
0 mmHg to 150 mmHg
3 mmHg to 50 mmHg
0 rpm to 150 rpm (Mainstream)
2 rpm to 150 rpm (Sidestream)
EtCO
2
FiCO
2
AwRR
± 2 mmHg, 0 mmHg to 40 mmHg
± 5% of reading, 41 mmHg to 70 mmHg
± 8% of reading, 71 mmHg to 100 mmHg
± 10% of reading, 101 mmHg to 150 mmHg
± 12% of reading, RR is over 80 rpm (sidestream)
There will be no degradation in performance due to
Respiration Rate. (mainstream)
± 1 rpm
Measure, standby
Flowrate
(50 ± 10) ml/min
0% to 100%
1%
16%
Pressure
User setup
0% to 20%
0.1%
0.0%
, AwRR
2
1 mmHg
1 mmHg
1 rpm
- 262 -