Patient Monitor User Manual
Message
BIS Artifact
RM
RM Comm Fail
RM Flow Module
Faulty
RM Flow Sensor Off
RM
Flow
Sensor
Error
CO
(RM) Comm.
2
Failed
CO
(RM) Occlude
2
Cause
Artifact,
such
as
generated by motion or eye
blinks.
RM module failure
RM module has malfunction
in the memory, barometric
pressure or hardware.
The flow sensor may be
disconnected
from
patient or the monitor.
Mismatch of sensor type
and patient type.
RM
module
failure
communication failure
The cannula is occluded.
Alarm Level
those
Low
High
High
the
Low
Low
or
High
High
- 66 -
Alarm Information
Action Taken
Attempt to identify
and
eliminate
artifact source.
Check if the module
is
properly
connected.
Stop
using
measuring
function
of
RM
module, and notify
biomedical engineer
or
manufacturer's
service staff.
Stop
using
measuring function
of RM module, and
notify
biomedical
engineer
or
Manufacturer's
service staff.
Check the sensor
connection.
Check consistency
of sensor type and
patient type.
Check if the module
is
properly
connected.
Stop
using
measuring
function
of
RM
module, and notify
biomedical engineer
or
manufacturer's
service staff.
Make sure the gas
exhaust works well