Prompts - EDAN elite V5 User Manual

Patient monitor
Table of Contents

Advertisement

Patient Monitor User Manual
Message
Battery1
Current
Too High
Battery2
Current
Too High
Battery1
Charge
Voltage Too High
Battery2
Charge
Voltage Too High
Battery
Error
iM20
Printer Unavailable
Insufficient storage
space
Read-only
storage
device
Audio Failed
Network Disconnect

5.3 Prompts

Message
V-Fib/V-Tach Off
Cause
Battery
1:
over-current
Battery
2:
over-current
Battery
1:
during charging
Battery
2:
during charging
in
Battery
communication failure
The selected printer is not
available.
Less than 10 M space is left
in the storage device.
The
storage
read-only.
Audio circuit connection is
abnormal, or loudspeaker
falls off.
In distributed alarm system,
the monitor's network is
disconnected.
Cause
V-Fib/V-Tach alarm is set to Off.
Alarm Level
discharge
Low
discharge
Low
over-voltage
Low
over-voltage
Low
module
Low
Low
Low
device
is
Low
High
Low
- 71 -
Alarm Information
Action Taken
Stop
using
the
battery and notify
the manufacturer's
service staff.
Check whether the
network connection
is in good condition
and
whether
the
printer
is
malfunctioning.
Delete some data in
the storage device
or
use
another
removable device.
Repair the storage
device or replace it
with a new one.
Stop
using
the
monitor and notify
the manufacturer's
service staff.
1) Check if the
network
cable
is
well connected.
2) Check if the
CMS is turned on.
3) Check if the IP of
bedside monitor and
CMS are on the
same
network
segment.

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Elite v6Elite v8

Table of Contents