Patient Monitor User Manual
ARR Alarms
Pacer
not
No QRS complex detected in 300 ms after a pace pulse.
Capture
Pacer
not
No pace pulse detected in 1.75 times RR interval after a QRS complex.
Pacing
Basic:
5 consecutive ventricular beats, and ventricular HR < 40 bpm.
Vent Brady
Advanced:
5 consecutive ventricular beats, and ventricular HR < 20 bpm.
Basic:
5 consecutive ventricular beats, and 40 bpm ≤ ventricular HR < 100 bpm.
Vent Rhythm
Advanced:
5 consecutive ventricular beats, and 20 bpm ≤ ventricular HR < 40 bpm.
The measurement value of PVCs is greater than high alarm limit that has
PVCs High
been set.
Applicable to Advanced ARR:
The duration of ventricular tachycardia rhythm ≥ the threshold value that has
Sustain VT
been set.
ExtremeTachy HR ≥ Extreme Tachycardia threshold value that has been set.
ExtremeBrady HR ≤ Extreme Bradycardia threshold value that has been set.
V-Tach
5 consecutive ventricular beats and ventricular HR ≥ 100 bpm.
Wide
QRS
Meet tachycardia conditions, and QRS wave width ≥ 160 ms.
Tachy
Non-Sustain
3 ≤ The number of consecutive ventricular beats < 5, and ventricular HR ≥
VT
100 bpm.
Atrial fibrillation alarm should meet below two conditions for 1 minute:
Afib
The RR interval of normal beats must be irregular, and it can be seen that the
obvious f or P waves do not exist.
Acc.
Vent
5 consecutive ventricular beats, and 40 bpm ≤ ventricular HR < 100 bpm.
Rhythm
No QRS is detected within the heartbeat pause threshold value that has been
Pause
set.
Pauses/min
The measurement value of Pause/min is greater than high alarm limit that has
High
been set.
The delayed ventricular beats detected in normal heartbeats occur more than
VEB
or equal to 3 times within 30 s.
Occurring Condition
- 102 -
Monitoring ECG