EDAN elite V5 User Manual page 88

Patient monitor
Table of Contents

Advertisement

Patient Monitor User Manual
Message
Zero OK
Pulsatile Pressure Zero Fail.
Pressure out of normal range,
Fail.
Sensor Off, Fail!
Invalid Time, Zero Fail.
Unable to Calibrate in Demo
Mode
Zeroing...
Please Press 'Calibrate'.
Calibration OK
Pulse
Pressure
Failed
Pressure out of range
Zeroing and Calibration Failed
Sensor Off, Fail.
Invalid Time, Calibration Fail.
Unable to Calibrate in Demo
Mode
Calibrating...
IBP alias collision
C.O. Lack Param
AG Self-Testing......
AG Span Calib In Progress
MultiGas Zero in Progress
AG Is Starting
AG Standby
AG Zero In Progress
AG Is Warming Up
AG Changing to Standby
Cause
IBP completes zeroing.
During the zeroing process, pressure fluctuation is
excessive.
During the zeroing process, pressure value is beyond the
zeroing range.
Perform zeroing when the sensor is off.
Time is not set up prior zeroing.
Perform zeroing in Demo Mode.
Zeroing is in progress.
Enter the Calibration menu, and Calibration is not
performed yet.
Calibration is completed.
During the Calibration process, pressure fluctuation is
Calibration
excessive.
During the Calibration process, pressure value is beyond
the Calibration range.
Zeroing is not performed prior calibration.
Perform calibration when the sensor is off.
Time is not set up prior calibration.
Perform calibration in Demo Mode.
Calibration is in progress.
The same IBP label appears.
Parameter is not configured for C.O. measurement.
AG module is performing a power-on self-test.
The calibration of AG module is in progress.
The zeroing of AG module is in progress.
Scio module is starting.
User sets Work Mode to Standby.
The zeroing of Scio module is in progress.
Scio module is warming up and is operating at reduced
accuracy
Work Mode is switching to Standby from Measure.
- 73 -