Patient Monitor User Manual
Message
ICG L1 Lead Off
ICG R1 Lead Off
ICG L2 or L3 Lead
Off
ICG R2 or R3 Lead
Off
ICG L4 Lead Off
ICG R4 Lead Off
ICG Comm Fail
NMT (TOF Watch)
NMT Stim. Signal
Missing
NMT
Acc.
Signal
Missing
Cause
The No.1 lead on the left is
off.
The No.1 lead on the right is
off.
The No.2 or No.3 lead on
the left is off.
The No.2 or No.3 lead on
the right is off.
The No.4 lead on the left is
off.
The No.4 lead on the right is
off.
The
communication
between the ICG module
and the monitor fails during
measuring.
During
stimulation,
electrode is moved, dropped
or broken. The monitor or
external
device
disconnected
with
stimulation cable.
During
stimulation,
sensor is moved, dropped or
broken. Or in calibration
mode, acc sensor connection
is not well or signal is too
weak.
Alarm Level
Low
Low
Low
Low
Low
Low
High
the
Medium
is
the
acc
Medium
- 68 -
Alarm Information
Action Taken
Make sure the No.1
lead on the left is
properly connected.
Make sure the No.1
lead on the right is
properly connected.
Make sure the No.2
or No.3 lead on the
left
is
properly
connected.
Make sure the No.2
or No.3 lead on the
right
is
properly
connected.
Make sure the No.4
lead on the left is
properly connected.
Make sure the No.4
lead on the right is
properly connected.
Unplug the module
and plug it again. If
the
problem
still
exists, contact your
service personnel.
Check the electrode
and cable to confirm
connection
to
patient, the monitor
and external device
cable
are
well
connected.