EDAN elite V5 User Manual page 72

Patient monitor
Table of Contents

Advertisement

Patient Monitor User Manual
Message
SpO
Low Perfusion
2
SpO
Noisy Signal
2
SpO
Light
2
Interference
NIBP
NIBP Comm Fail
NIBP Leak
NIBP
Excessive
Pressure
NIBP Init Pressure
High
Cause
The pulse signal is too
weak or the perfusion of
the measurement site is too
low. The SpO
value and
2
PR
value
might
inaccurate then.
There is interference with
SpO
measurement signals
2
due to patient movement,
ambient
light,
electrical
interference or else.
Ambient light around the
sensor is too strong.
NIBP module failure or
communication failure
NIBP pump, valve, cuff or
tube has a leakage.
Pressure has exceeded the
specified upper safety limit.
The initial pressure is too
high during measuring
Alarm Level
Low
be
Low
(EDAN
SpO
module)
2
Low
High
Low
Low
Low
- 57 -
Alarm Information
Action Taken
Reconnect the SpO
2
sensor and change
the
measurement
site.
If
problem
exists, please notify
biomedical engineer
or
manufacturer's
service staff.
Check the condition
of patient and avoid
patient movement;
make sure the cable
is well connected.
Reduce interference
of the ambient light
and avoid sensor's
exposure to strong
light.
Stop
using
measuring function
of NIBP module,
and
notify
biomedical engineer
or
manufacturer's
service staff.
Check
the
connections and the
wrapped cuff to see
whether they are all
prepared well.
Measure again, if
failure persists, stop
measuring function
of NIBP module
and
notify
biomedical engineer
or
manufacturer's
service staff.