Patient Monitor User Manual
Click on the Anes/Vent Interface, open the setup menu, the user can:
•
Select the parameter to be displayed on the window in the Parameter list.
•
Select a suitable ruler for the waveform from the options TopRuler , MidRuler and
BotRuler . MidRuler is not available when anesthesia device is connected.
•
Select an appropriate sweep for the waveform in the Sweep list. The bigger the value is, the
wider the waveform is.
•
Choose Mode and set it to Curve or Filled from the pop-up list.
•
Select Color Setup to make color changes on parameter and waveform.
•
Select Default to restore factory default for all setups in this menu.
•
Select Enter Full-param Interface to display all the sub-parameters' measurement values,
without waveforms displayed.
•
Select Respiratory Loop to enter respiratory loop interface (only applicable to Maquet
ventilator).
•
View the Unit and Baro Pess .
NOTE:
1
When the ventilator is working, it must be ventilated. If not ventilated (air source is
disconnected), there may be a risk.
2
The monitor does not provide alarms from the ventilator itself.
3
When switching between different devices, the trend data of the previous device is
cleared.
21.6.2.1 Respiratory Loop Interface
Normally, click on the Anes/Vent window to open the setup menu, select Loop in the Parameter
list, and user can enter respiratory loop interface by clicking Large Loop .
Respiratory loops can indicate a fault in the airway tubing and help physicians to detect
respiratory problems of patients. The two types of loops are available in real time:
•
F-V (flow-volume) loops: it illustrates the dynamic relation between flow and volume during
respiration and provides information about condition of the airway tubing.
•
P-V (pressure-volume) loops: it reflects the dynamic relation between pressure and volume
as well as compliance of the respiratory system.
- 190 -