Patient Monitor User Manual
ARR Alarm
PAC Bigeminy
PAC Trigeminy
Pauses/min High
Sustain VT
ExtremeBrady
Low Voltage(Limb)
8.9.2.4 ARR Selflearning
Pick this item ARR Selflearn to start a learning procedure, and ECG ARR Learning is
displayed on the screen.
The ARR selflearning will start automatically in the following status:
Changing patient type or lead type;
Connecting or Switching calculation leads;
Changing pacemaker status;
Exiting DEMO or Standby mode;
Admitting a patient;
Switching calibration mode into normal measurement mode;
Switching the ECG parameter on;
Changing-over between basic ARR and advanced ARR.
NOTE:
1
During the relearning phase of the algorithm, arrhythmia detection may not be
available. So you should closely monitor the patient condition during and for several
minutes after the learning phase to allow the algorithm to reach optimal detection
performance.
2
Take care to initiate ARR selflearning only during periods of predominantly normal
rhythm and when ECG signal is relatively noise-free. If ARR selflearning takes place
during arrhythmia, the ectopics may be incorrectly learned as normal QRS complex.
This may result in missed detection of subsequent events of arrhythmia.
3
If ARR selflearning is performed during ventricular rhythm, ventricular heartbeats
may be erroneously identified as normal QRS complexes. This may lead to missed
ventricular tachycardia and ventricular fibrillation events.
Due to this reason, you should:
1) Take care that ARR selflearning may start automatically;
2) Response to lead off information;
3) Always check the correctness of arrhythmia alarm.
Range
3/min to 50/min
1/min to 20/min
15 s to 45 s
15 bpm to 60 bpm
0.3 mV to 0.8 mV
- 106 -