EDAN elite V5 User Manual page 297

Patient monitor
Table of Contents

Advertisement

Patient Monitor User Manual
Electrostatic
discharge (ESD)
IEC/EN 61000-4-2
Electrical fast
transient/burst
IEC/EN 61000-4-4
Surge
IEC/EN 61000-4-5
Power frequency
(50 Hz /60 Hz)
magnetic field
IEC/EN 61000-4-8
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC/EN
61000-4-11
NOTE
U
is the a.c. mains voltage prior to application of the test level.
T
± 8 kV contact
± 15 kV air
± 2 kV for power supply
lines
± 1 kV for line to line
± 2 kV for line to
ground
30 A/m
0 % U
0.5 cycle
T;
At 0°, 45°, 90°, 135°,
180°, 225°, 270° and
315°
0 % U
; 1 cycle
T
and
70 % U
; 25/30 cycles )
T
Single phase: at 0°
0 % U
; 250/300 cycle
T
± 8 kV contact
± 15 kV air
± 2 kV for power
supply lines
± 1 kV for line to
line
± 2 kV for line to
ground
30 A/m
0 % U
0.5 cycle
T;
At 0°, 45°, 90°,
135°, 180°, 225°,
270° and 315°
0 % U
; 1 cycle
T
and
70 % U
; 25/30
T
cycles )
Single phase: at 0°
0 % U
; 250/300
T
cycle
- 282 -
EMC Information
Floors should be wood,
concrete or ceramic tile.
If floor are covered with
synthetic material, the
relative humidity should
be at least 30%.
Mains power quality
should be that of a
typical commercial or
hospital environment.
Mains power quality
should be that of a
typical commercial or
hospital environment.
Power frequency
magnetic fields should
be at levels characteristic
of a typical location in a
typical commercial or
hospital environment.
Mains power quality
should be that of a
typical commercial or
hospital environment. If
the user of elite V5 V6
V8 requires continued
operation during power
mains interruptions, it is
recommended that elite
V5 V6 V8 be powered
from an uninterruptible
power supply or a
battery.

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Elite v6Elite v8

Table of Contents