Note 1: Negligible interference, effect included in the specification "Accuracy, all conditions"
above.
Note 2: Negligible interference with N
specification "Accuracy, all conditions" above.
Note 3: Interference at indicated gas level. For example, 50 vol% Helium typically decreases the
CO
readings by 6%. This means that if measuring on a mixture containing 5.0vol% CO
2
50vol% Helium, the actual measured CO
4.7vol% CO
.
2
Note 4: According to the EN ISO 80601-2-55:2011 standard.
Note 5: In addition to the EN ISO 80601-2-55:2011 standard.
Dräger Minimodule
Method
Barometric
Compensation
Gas Sampling Rate
Maximum time until water trap
requires draining
Total System Response Time
Drift Compensation (zeroing)
Zeroing Duration
Cross Sensitivity
☆ O
2
☆ Range
1
☆ Accuracy
Rise Time (t10...90)
Time to Specified Accuracy
☆ CO
2
_ 1)
0.02 vol%
_ 2)
100 vol%
O / O
2
2
Sidestream gas measurement
Infrared measurement: CO
Paramagnetic measurement: O
Pressure
Automated compensation
200 mL/min ±20 mL/min
41 hrs (sample gas under BTPS conditions, ambient air
23 °C)
< 3 s
Automated cyclical zeroing,
once per day (in error-free operation)
< 20 s
None concerning alcohol (< 3000 ppm blood conc.),
acetone (< 1000 ppm), methane, water vapor, NO, and CO
(0 to 100) Vol%
± (2.5 Vol% + 2.5 % rel.)
4
< 500 ms
3
< 450 s
_ 1)
_ 1)
concentrations correctly set, effect included in the
2
concentration will typically be (1-0.06) * 5.0vol% =
, N
2
- 268 -
Product Specifications
_ 1)
_ 1)
O, anesthetic agents
2
2
and
2