Patient Monitor User Manual
AwRR 1 rpm
EtCO
☆ Accuracy
AwRR ± 1 rpm
Drift of
Measure
Meets the requirements of the measure accuracy
Accuracy
Sample
Gas
70 ml/min or 100 ml/min(default), accuracy: ± 15 ml/min
Flowrate
Warm-upTime Display reading within 20 s; reach to the designed accuracy within 2 minutes.
< 400 ms (with 2 m gas sampling tube, sample gas flowrate: 100 ml/min)
Rise Time
< 500 ms (with 2 m gas sampling tube, sample gas flowrate: 70 ml/min)
< 4 s (with 2 m gas sampling tube, sample gas flowrate: 100 ml/min)
Response
Time
< 4 s (with 2 m gas sampling tube, sample gas flowrate: 70 ml/min)
Work Mode
Standby, measure
Range: 0% to 100%
O
2
Resolution: 1%
Compensation
Default: 16%
Range: 0% to 100%
N
O
2
Resolution: 1%
Compensation
Default: 0%
Range: 0% to 20%
AG
Resolution: 0.1%
Compensation
Default: 0%
± 2 mmHg, 0
mmHg
to
40
mmHg
± 5% of reading,
41 mmHg to 70
mmHg
± 8% of reading,
71 mmHg to 100
2
mmHg
± 10% of reading,
101 mmHg to 150
mmHg
± 12% of reading
or ± 4 mmHg,
whichever
greater
Respiratory
rate ≤ 60 rpm
Respiratory
rate > 60 rpm
is
- 260 -
Product Specifications
Typical conditions:
Ambient temperature: (25 ± 3)
° C
Barometric pressure: (760 ±
10) mmHg
Balance gas: N
2
Sample gas flowrate: 100
ml/min
All conditions