EDAN elite V5 User Manual page 284

Patient monitor
Table of Contents

Advertisement

Patient Monitor User Manual
☆ Range
1
☆ Accuracy
Rise Time (t10...90)
Time to availability
Time to Specified Accuracy
☆ N
O
2
☆ Range
1
☆ Accuracy
Rise Time (t10...90)
Time to Specified Accuracy
☆ Anesthetic Gases Range
☆ Halothane
☆ Isoflurane
☆ Enflurane
☆ Sevoflurane
☆ Desflurane
1
☆ Accuracy
Rise Time (t10...90)
Time to Specified Accuracy
Automatic Detection
Primary Gas
Secondary Gas
☆ Respiratory Rate
☆ Range
☆ Accuracy
Resolution
(0 to 13.6) Vol%
± (0.43 Vol% + 8% rel.)
4
< 350 ms
2
< 60 s
3
< 450 s
(0 to 100) Vol%
± (2 Vol% + 8% rel.)
4
< 350 ms
3
< 450 s
(0 to 8.5) Vol%
(0 to 8.5) Vol%
(0 to 10) Vol%
(0 to 10) Vol%
(0 to 20) Vol%
± (0.2 Vol% + 15 % rel.)
4
< 450 ms
3
< 450 s
At the latest at 0.3 Vol%
At the latest at 0.4 Vol%
With a Desflurane concentration greater than 4 Vol%,
mixture detection occurs at the latest when the concentration
of the second anesthetic gas rises above 10% of the
Desflurane concentration.
0/min to 100/min (Respiratory rate is derived from the CO
value)
0/min to 60/min: ± 1 /min
> 60 /min: not specified
1 /min
- 269 -
Product Specifications