Bis - EDAN elite V5 User Manual

Patient monitor
Table of Contents

Advertisement

Patient Monitor User Manual
Note 2: IRMA AX+ does not measure O
Note 3: Interference at indicated gas level. For example, 50 vol% Helium typically decreases the
CO
readings by 6%. This means that if measuring on a mixture containing 5.0vol% CO
2
50vol% Helium, the measured CO
CO
.
2
Note 4: According to the EN ISO 80601-2-55:2011 standard.
Note 5: In addition to the EN ISO 80601-2-55:2011 standard.
A.21 BIS
Complies with IEC 60601-2-26: 2012.
Technique
☆ Measure
Parameters
Sweep Speed
Wave Scale
BIS Trend
Smoothing Rate
Noise
(EEG
Waveform)
EEG Bandwidth 0.25 Hz ~ 50 Hz
☆ BIS Alarm
Range
A.22 RM
Complies with ISO 80601-2-55: 2011.
Measure Parameters
Sensor Zero
Frequency Response > 10 Hz
concentration will typically be (1-0.06) * 5.0 vol% = 4.7vol%
2
Bispectral index, power spectrum analysis
Primary Parameter
Secondary Parameters
6.25 mm/s, 12.5 mm/s, 25 mm/s, 50 mm/s
50 μv, 100 μv, 200 μv, 500 μv
Length of BIS trend: 6 min, 12 min, 30 min, 60 min
10 s, 15 s, 30 s
< 0.3 μV (0.25 Hz ~ 50 Hz)
0 ~ 100
Flow, Tidal Volume, Airway Pressure, Respiration Rate
Typically 2 seconds. Maximum zero interval is 10 minutes for an adult and
pediatric sensor, and 3 minutes for a neonatal sensor.
.
2

BIS

SQI
SR
EMG
SEF
TP
BC (only applicable to
BIS™ Extend Sensor)
- 273 -
Product Specifications
0 to 100
0% to 100%
0% to 100%
30 dB to 80 dB
0.5 Hz to 30.0 Hz
40 dB to 100 dB
0 to 30
and
2