EDAN elite V5 User Manual page 276

Patient monitor
Table of Contents

Advertisement

Patient Monitor User Manual
Humidity
Compensation
ATPD (default), BTPS
Method
Barometric
Automatic (The change of barometric pressure will not add additional errors
Pressure
to the measurement values.)
Compensation
Zero
Support
Calibration
Calibration
Support
EtCO
☆ Alarm
10 s, 15 s, 20 s, 25 s, 30 s, 35 s, 40 s, 60s; default value is 20 s.
Apnea
Alarm Delay
Data
Sample
100 Hz
Rate
AwRR ≤
mentioned above;
AwRR > 80 rpm, EtCO
AwRR > 120 rpm, EtCO
EtCO
2
AwRR ≤
1
Change
mentioned above;
AwRR > 60 rpm, EtCO
AwRR > 90 rpm, EtCO
AwRR > 120rpm, EtCO
Note 1: Use a test device equivalent to EN ISO 80601-2-55 fig 201.101 to measure at 1:2 I/E
ratio. Respiration rate accuracy is determined by frequency of device, and end-tidal reading
change refers to the nominal value.
Interfering Gas Effects:
Gas
Nitrous oxide
Halothane
Enflurane
Isoflurane
Sevoflurane
Desflurane
Respironics Module
Applicable Patient Type
Technique
, FiCO
, AwRR
2
2
80rpm,
meet
descends 8%;
2
2
60rpm,
meet
descends 8%;
2
descends 10%;
2
descends 15%;
2
Gas Level (%)
60
4
5
5
5
15
Adult, pediatric and neonatal patients
Infra-red Absorption Technique
the
accuracy
descends 10%
the
accuracy
Quantitative Effect/Comments
The interfering gas will have no effect on the
measurement value if compensation of O
anesthetic agents has been correctly set.
- 261 -
Product Specifications
with 2 m gas sampling tube,
sample gas flowrate: 100
ml/min)
with 2 m gas sampling tube,
sample
gas
flowrate:
ml/min)
70
, N
O,
2
2