Patient Monitor User Manual
2
When iM20 (as sub-monitor) is connected to VX series monitor, if iM20 supports
Advanced ARR, it will synchronize with elite VX series monitor; if not, Basic ARR will
be adopted.
3
When XM module or iM20 has been inserted into VX series monitor, user unplugs the
inserted one from the monitor and reinserts another one, if the ARR analyses of
newly inserted one is inconsistent with that of the previous one, the monitor will pop
up a prompt to notify user that ARR analysis is switched to Basic or Advanced.
4
The measured PVCs and Pauses/min will be displayed in main interface.
Pauses/min is only applicable to Advanced ARR.
ARR Alarms
Applicable to both Basic ARR and Advanced ARR
Asystole
No QRS is detected for 4 consecutive seconds
4 consecutive seconds' fibrillation wave occurs; Or 5 consecutive ventricular
V-Fib/V-Tach
beats, and ventricular HR ≥100 bpm.
Run PVCs
3 ≤ the number of consecutive PVCs < 5
Couplet
2 consecutive PVCs
PVC
A dominant rhythm of N, V, N, V (N = supraventricular beat, V =
Bigeminy
ventricular beat) was detected.
PVC
A dominant rhythm of N, N, V, N, N,V
Trigeminy
A type of single PVC under the condition that HR<100, R-R interval is less
R on T
than 1/3 the average interval, followed by a compensating pause of 1.25X the
average R-R interval (the next R wave advances onto the previous T wave).
Single PVC detected in normal heartbeats, and the number of consecutive
PVC
single PVC ≥ 4 within 30 s.
Adult: RR interval for 5 consecutive QRS complex ≤ 0.5 s.
Tachy
Pediatric/neonatal: RR interval for 5 consecutive QRS complex ≤ 0.375 s.
Adult: RR interval for 5 consecutive QRS complex ≥ 1.5 s.
Brady
Pediatric/neonatal: RR interval for 5 consecutive QRS complex ≥ 1 s.
Basic:
If HR < 120 bpm, no beats are detected for 1.75 times average RR interval;
or if HR
Missed Beat
wave is detected within 3 s or longer.
Advanced:
If HR < 120 bpm, no beats are detected for 1.75 times average RR interval;
or if HR
Irr Rhythm
Consistently irregular heart rhythm
Occurring Condition
120 bpm, no beats are detected for one second; or no valid QRS
120 bpm, no beats are detected for one second;
- 101 -
Monitoring ECG