Patient Monitor User Manual
Message
CO
(RM)
Check
2
Adapter
CO
(RM)
Sensor
2
Faulty
CO
(RM)
Sensor
2
Over Temp
CO
(RM)
Zero
2
Required
CO
(RM) Out Of
2
Range
CO
(RM)
Sensor
2
Off
ICG
ICG Sensor Off
ICG
Need
Input
Param
ICG Input Param
Error
Cause
The cannula is off or
disconnected.
CO
module failure
2
CO
measure value exceeds
2
the measure range of the
monitor.
Zero calibration failure
The
CO
concentration
2
exceeds the accuracy range
of RM module.
The
sensor
may
disconnected
from
patient or the monitor.
1) The ICG sensor is
disconnected
from
module
2) Bad connection
The required patient data,
such as height and weight,
has not been input into the
monitor.
The input patient data is
invalid.
Alarm Level
Low
High
High
Low
High
be
the
Low
the
Low
Low
Low
- 67 -
Alarm Information
Action Taken
Check whether the
adapter is properly
connected
or
replace the adapter.
Stop
using
measuring function
of
CO
module,
2
notify
biomedical
engineer.
Disconnect
the
sampling cannula or
adapter
from
the
airway; initiate the
zeroing
before
making sure that no
expired air is inside
the
sampling
cannula and adapter.
Reduce
CO
2
concentration.
Check the sensor
connection.
Reconnect the ICG
sensor.
Input patient data.
Input valid patient
data.