Patient Monitor User Manual
Data Sample Rate
Respiration Rate
Fi and ET
Measurement Range
Resolution
Gas
Accuracy-
CO
2
Standard
Conditions
N
O
2
HAL, ENF, ISO
SEV
DES
O
2
20 data frames, 420 bytes, per second
0 to 150 ± 1 breaths/min
Fi and ET are displayed after one breath and have a continuously
updated breath average.
ET will typically decrease below nominal value (ET
respiration rate (RR) exceeds the RR threshold (RR
following formulas:
CO
2
N
O, O
, DES, ENF, ISO , SEV
2
2
HAL
NOTE: Measured at I/E ratio 1:1 using breath simualtor according to
EN ISO 80601-2-55 fig. 201.101
CO
: (0 to 25) vol%
2
O
: (0 to 100) vol%
2
N
O: (0 to 100) vol%
2
HAL, ENF, ISO, SEV, DES: (0-25) vol%
AwRR: 0 rpm to 150 rpm
CO
: 0.1%
2
HAL, ENF, ISO, SEV, DES: 0.1%
N
O: 1%
2
O
:1%
2
AwRR: 1 rpm
Range
(0 to 15) vol%
(15 to 25) vol%
(0 to 100) vol%
(0 to 8) vol %
(8 to 25) vol %
(0 to 10) vol %
(10 to 25) vol %
(0 to 22) vol %
(22 to 25) vol %
(0 to 100) vol %
- 266 -
Product Specifications
) when
nom
) according to the
th
Accuracy
± (0.2 vol% + 2% of reading)
Unspecified
± (2 vol% + 2% of reading)
± (0.15 vol% + 5% of reading)
Unspecified
± (0.15 vol% + 5% of reading)
Unspecified
± (0.15 vol% + 5% of reading)
Unspecified
± (1 vol% + 2% of reading)