ST value
Range
Accuracy
Resolution
HR Averaging Method
Method 1
Method 2
Range of Sinus and SV Rhythm
Tachy
Normal
Brady
Range of Ventricular Rhythm
V-Tach
Vent Rhythm
Vent Brady
Maximum Start-up Alarm Time for Tachycardia
-2.0 mV to +2.0 mV
-0.8 mV to +0.8 mV: ±0.02 mV or 10%, whichever is greater.
Beyond this range: not specified.
0.01 mV
Heart rate is computed by excluding the minimum and maximum
values from the 12 most recent RR intervals and averaging the
residual 10 RR intervals.
If each of three consecutive RR intervals is greater than 1200 ms, then
the four most recent RR intervals are averaged to compute the HR.
Adult: RR interval for 5 consecutive QRS complex ≤ 0.5 s.
Pediatric/neonatal: RR interval for 5 consecutive QRS complex ≤
0.375 s.
Adult: 0.5 s < RR interval for 5 consecutive QRS complex < 1.5 s.
Pediatric/neonatal: 0.375 s < RR interval for 5 consecutive QRS
complex < 1 s.
Adult: RR interval for 5 consecutive QRS complex ≥ 1.5 s.
Pediatric/neonatal: RR interval for 5 consecutive QRS complex ≥ 1 s.
5 consecutive ventricular beats and ventricular HR ≥ 100 bpm.
Basic:
5 consecutive ventricular beats, and 40 bpm ≤ ventricular HR < 100
bpm.
Advanced:
5 consecutive ventricular beats, and 20 bpm ≤ ventricular HR < 40
bpm.
Basic:
5 consecutive ventricular beats, and ventricular HR < 40 bpm.
Advanced:
5 consecutive ventricular beats, and ventricular HR < 20 bpm.