3.12 Resolving IBP Label Conflicts

Each label must be unique and can only be assigned once. The measurement labels are stored in
the measurement modules. If you try to use two IBP channels that have identical labels, this
causes a label conflict in the monitor.
For example, an IBP channel (channel A) has already been loaded and the label Art is used for
channel A. Then another IBP channel (channel B) is loaded and the label Art is also used for
channel B. In this case, a label conflict will be triggered. A prompt indicating IBP label conflict
will appear on the left of the screen. Additionally, at the corresponding measurements area, a
label flickers to indicate a label conflict. The flickering label is the default one assigned by the
system.
The IBP channel with a label conflict will not provide any measurement data; besides, the
functions of setup, zeroing and calibrating are unavailable. To resolve the label conflict, you have
to change the conflicting label into a non-conflicting one via the following method:
1.
Select the IBP channel with a label conflict on the screen and open the Options menu.
2.
Choose another label among the options from the Alias pull-down list to resolve the label
conflict.
- 33 -