8.5.4.2 Electrode Placement for 5-Lead
Take the American standard for example; see the following figure:
RA placement: directly below the clavicle and near the right shoulder.
LA placement: directly below the clavicle and near the left shoulder.
RL placement: on the right hypogastrium.
LL placement: on the left hypogastrium.
V placement: on the chest, the position depends on your required lead selection.
NOTE:
To ensure the patient safety, all leads must be attached to the patient.
For 5-lead, attach the V electrode to one of the indicated positions as below:
V1
On the 4th intercostal space at the right sterna margin.
V2
On the 4th intercostal space at the left sterna margin.
V3
Midway between V2 and V4 electrodes.
V4
On the 5th intercostal space at the left clavicular line.
V5
On the left anterior axillary line, horizontal with V4 electrode.
V6
On the left middle axillary line, horizontal with V4 electrode.
V3R-V6R
On the right side of the chest in positions corresponding to those on the left.
VE
Over the xiphoid position.
V7
On the 5th intercostal space at the left posterior axillary line of back.
V7R
On the 5th intercostal space at the right posterior axillary line of back.
Electrode Placement for 5-Lead
- 92 -
Monitoring ECG