Patient Monitor User Manual
Balance Gas Compensation
Stability
Short Term Drift
Long Term Drift
Zero Calibration
☆ Alarm Type
☆ Apnea Alarm Delay
Data Sample Rate
CO
Rise Time/Response Time
2
(mainstream)
Sensor
Response
(sidestream)
Interfering Gas and Vapor Effects on EtCO
Gas or Vapor
Nitrous oxide
Halothane
Enflurane
Isoflurane
Sevoflurane
Xenon
Helium
Desflurane
Room air, N
Drift over 4 hours < 0.8 mmHg
120 hours
Support
EtCO
, FiCO
2
10 s, 15 s, 20 s, 25 s, 30 s, 35 s, 40 s; default value is 20 s.
100 Hz
Less than 60 ms
Time
< 3 seconds, including transport time and rise time
Measurement Values:
2
Gas Level (%)
Quantitative Effect/Comments
60
Dry and Saturated Gas
4
(0 ~ 40) mmHg: ±1 mmHg additional error
5
(41 ~ 70) mmHg: ±2.5% additional error
5
(71 ~ 100) mmHg: ±4% additional error
5
(101 ~ 150) mmHg: ±5% additional error
80
*Additional worst case error when compensation
for P
50
correctly selected for the actual fractional gas
15
constituents present.
Desflurane:
The presence of desflurane in the exhaled breath at
concentrations greater than 5% will positively bias
Carbon Dioxide values by up to an additional 3
mmHg at 38 mmHg.
Xenon:
The presence of Xenon in the exhaled breath will
negatively bias Carbon Dioxide values by up to an
additional 5 mmHg at 38 mmHg.
- 263 -
O, helium
2
, AwRR
2
, O
, N
O, anesthetic agents, or helium is
B
2
2
Product Specifications