SHOCK HAZARD - Don't connect electrical equipment, which has not been supplied
as a part of the system, to the multiple portable socket-outlet supplying the system.
43
SHOCK HAZARD - Don't connect electrical equipment, which has been supplied as a
part of the system, directly to the wall outlet when the non-medical equipment is
intended to be supplied by a multiple portable socket-outlet with an isolation
transformer.
44
To protect the monitor from damage during defibrillation, for accurate measurement
information and to protect against noise and other interference, use only accessories
specified by EDAN.
45
This equipment is not intended for home usage.
46
Operation of the equipment exceeding specified physiological signal or the
operational specification may cause inaccurate results.
47
When the monitor is used with HF surgical equipment, the transducer and the cables
must be avoided from conductive connection to the HF equipment. This is to protect
against burns to the patient.
48
No modification of this equipment is allowed without authorization of the manufacturer.
If this equipment is modified, appropriate inspection and testing must be conducted to
ensure continued safe operation.
49
The monitor is suitable for use in the presence of electrosurgery. When the monitor is
used with HF surgical equipment, user (doctor or nurse) should be cautious about
patient safety.
50
Make sure networking function is used in a secure network environment.
51
The device must be connected to the ground to avoid signal interference.
52
Portable RF communications equipment (including peripherals such as antenna
cables and external antennas) should be used no closer than 30 cm (12 inches) to
any part of the monitor, including cables specified by the manufacturer. Otherwise,
degradation of the performance of this equipment could result.
1
Electromagnetic Interference - Ensure that the environment in which the patient
monitor is installed is not subject to any sources of strong electromagnetic
interference, such as radio transmitters, mobile telephones, microwaves, etc.
2
Keep the environment clean. Avoid vibration. Keep it far away from corrosive
medicine, dust area, high temperature and humid environment.
3
Do not immerse transducers in liquid. When using solutions, use sterile wipes to avoid
pouring fluids directly on the transducer.
WARNING
CAUTION
- 5 -