EDAN elite V5 User Manual page 63

Patient monitor
Table of Contents

Advertisement

Patient Monitor User Manual
Message
Brady
Missed Beat
Irr Rhythm
Pacer
not
Capture
Pacer not Pacing
Vent Brady
Vent Rhythm
Sustain VT
ExtremeTachy
ExtremeBrady
V-Tach
Cause
Adult: RR interval for 5 consecutive QRS complex ≥
1.5 s.
Pediatric/neonatal: RR interval for 5 consecutive QRS
complex ≥ 1 s.
Basic:
If HR < 120 bpm, no beats are detected for 1.75 times
average RR interval; or if HR
detected for one second; or no valid QRS wave is
detected within 3 s or longer.
Advanced:
If HR < 120 bpm, no beats are detected for 1.75 times
average RR interval; or if HR
detected for one second.
Consistently irregular heart rhythm
No QRS complex detected in 300 ms after a pace
pulse.
No pace pulse detected in 1.75 times RR interval after
a QRS complex.
Basic:
5 consecutive ventricular beats, and ventricular HR <
40 bpm.
Advanced:
5 consecutive ventricular beats, and ventricular HR <
20 bpm.
Basic:
5 consecutive ventricular beats, and 40 bpm ≤
ventricular HR < 100 bpm.
Advanced:
5 consecutive ventricular beats, and 20 bpm ≤
ventricular HR < 40 bpm.
The duration of ventricular tachycardia rhythm ≥ the
threshold value that has been set.
HR ≥ Extreme Tachycardia threshold value that has
been set.
HR ≤ Extreme Bradycardia threshold value that has
been set.
5 consecutive ventricular beats and ventricular HR ≥
100 bpm.
120 bpm, no beats are
120 bpm, no beats are
- 48 -
Alarm Information
Alarm Level
User-selectable
User-selectable
User-selectable
User-selectable
User-selectable
High
User-selectable
High
High
High
High