Patient Monitor User Manual
Respiration Rate
Fi and ET
Data Update Period
Measurement Range
Resolution
☆ Accuracy- Standard
Conditions
☆ Accuracy- All
Conditions
(0-150) ± 1 bpm. The respiration rate is displayed after three breaths
and the average value is updated every breath.
IRMA AX+: CO
, N
2
ISO, SEV, DES)
Fi and ET are displayed after one breath and have a continually
updated breath average.
ET-values for anathesthetic agents and N
typically decrease below nominal value when respiration rate
exceeds 80 bpm according to the formula ET = 80*ET
at I/E ratio 1:1 using breath simualtor according to EN ISO
80601-2-55 fig. 201.101).
EtCO
will be within specification for all respiration rates up to 150
2
bpm (IRMA AX+).
1 s
CO
: (0 to 25) vol%
2
N
O: (0 to 100) vol%
2
HAL, ENF, ISO, SEV, DES: (0-25) vol%
AwRR: 0 rpm to 150 rpm
CO
: 0.1%
2
HAL, ENF, ISO, SEV, DES: 0.1%
N
O: 1%
2
AwRR: 1 rpm
Gas
Range
CO
(0 to 15) vol%
2
N
O
(0 to 100) vol%
2
HAL
(0 to 8) vol%
ISO
ENF
SEV
(0 to 10) vol%
DES
(0 to 22) vol%
Gas
Accuracy
CO
± (0.3 kPa + 4% of reading)
2
N
O
± (2 kPa + 5% of reading)
2
Agents
± (0.2 kPa + 10% of reading)
- 271 -
O, primary and secondary agents (HAL, ENF,
2
Accuracy
± (0.2 vol% + 2% of reading)
± (2 vol% + 2% of reading)
± (0.15 vol% + 5% of reading)
± (0.15 vol% + 5% of reading)
± (0.15 vol% + 5% of reading)
Product Specifications
O (IRMA AX+) will
2
/RR (tested
nom