Maximum Cuff Pressure
Maximum
Measuring
Period
Pre-inflation Pressure
Adult/ Pediatric Mode
Neonatal Mode
Dual Independent Channel Overpressure Protection
Adult/Pediatric
Neonatal
SunTech Module
Method
Mode
Measuring Interval in AUTO Mode
☆ Measuring Type
☆ Measuring Range
☆ Adult Mode
☆ Pediatric Mode
☆ Neonatal Mode
☆ Alarm Type
Pressure Resolution
☆ Maximum mean error
☆ Maximum standard deviation
Maximum measuring period
300 mmHg
Less
than
160 s
Default: 180 mmHg
Range: 120/140/150/160/180/200/220/240/260/280 mmHg
Default: 120 mmHg
Range: 80/100/120/140 mmHg
< 300 mmHg
< 150 mmHg
Oscillometric
Manual, Auto, Continuous
1/2/3/4/5/10/15/30/60/90/120/240/480 (unit: minute)
SYS, DIA, MAP
SYS: 40 mmHg ~ 260 mmHg
DIA: 20 mmHg ~ 200 mmHg
MAP: 26 mmHg ~ 220 mmHg
SYS: 40 mmHg ~ 230 mmHg
DIA: 20 mmHg ~ 160 mmHg
MAP: 26 mmHg ~ 183 mmHg
SYS: 40 mmHg ~ 130 mmHg
DIA: 20 mmHg ~ 100 mmHg
MAP: 26 mmHg ~ 110 mmHg
SYS, DIA, MAP
1 mmHg
±5 mmHg
8 mmHg
- 255 -
330 mmHg
150 mmHg
Less than 180 s
Less than 80 s
Product Specifications
165 mmHg
Less than 90
s

Hide quick links: