Patient Monitor User Manual
Message
NIBP Aux Excessive
Pressure
NIBP Time Out
NIBP
Self
Test
Error
NIBP
Cuff
Type
Error
NIBP
System
Pressure
Abnormality
NIBP
System
Failure
NIBP Weak Signal
NIBP
Range
Exceeded
NIBP Loose Cuff
Cause
Pressure has exceeded the
second
safety
limit
specified.
Measuring
time
exceeded the specified time.
Sensor or other hardware
errors.
The cuff type used isn't
consistent with the patient
type.
Atmospheric pressure or
system
pressure
abnormal. The valve is
occluded so that deflation is
failed.
NIBP is not calibrated.
Cuff is too loose or patient
pulse is too weak.
Maybe the patient blood
pressure value is beyond the
measurement range.
Cuff
is
not
wrapped or no cuff is
connected.
Alarm Level
as
High
has
Low
Low
Low
is
Low
High
Low
High
properly
Low
- 58 -
Alarm Information
Action Taken
Notify
biomedical
engineer
or
manufacturer's
service staff.
Measure again or
use other measuring
method.
If failure persists,
stop
using
measuring function
of NIBP module
and
notify
biomedical engineer
or
manufacturer's
service staff.
Confirm the patient
type and change the
cuff.
Check whether the
airway is occluded
or pressure sensor
works properly. If
the
problem
still
exists, contact your
service personnel.
Contact
your
service personnel.
Use other methods
to measure blood
pressure.
Use other methods
to measure blood
pressure.
Properly wrap the
cuff.