Message
IBP Catheter Off
IBP Sensor Err
YY Comm Fail (YY
stands for the IBP
label name: Art, PA,
CVP,
RAP,
LAP,
ICP, P1 and P2)
C.O.
C.O. Comm Fail
C.O. TI Sensor Off
C.O. TB Sensor Off
C.O. TEMP Out Of
Range
AG
AG Baro Press Out
Of Range
AG Mixed Agents
MAC < 3
AG Zero Required
Cause
IBP catheter falls off due to
patient movement.
Malfunction in the IBP
sensor or in the extension
cable.
IBP
module
failure
communication failure
C.O.
module
failure
communication failure
C.O.
TI
sensor
connected
C.O.
TB
sensor
connected
TI/TB measuring value is
beyond measuring range.
The
barometric
exceeds
the
specified
working
barometric
pressure range.
Two types of anesthetic
agents are present in the gas
mixture,
and
concentration is low.
Zeroing of AG module is
required.
Alarm Level
High
Medium
or
High
or
High
not
Low
not
Low
High
pressure
High
Low
the
Medium
- 60 -
Alarm Information
Action Taken
Check the catheter
connection
and
reconnect it.
Replace
the
IBP
sensor
or
the
extension cable.
Stop
measuring
function
of
IBP
module, and notify
biomedical engineer
or
Manufacturer's
service staff.
Stop measuring of
C.O.
module,
or
notify
biomedical
engineer
or
Manufacturer's
service staff.
Insert
injective
temperature sensor.
Insert TB sensor.
Please check TI/TB
sensor.
Make sure the AG
module
is
used
within the specified
barometric pressure
range.
Adjust
the
concentration of the
anesthetic agents if
necessary.
Perform
zero
calibration.