Patient Monitor User Manual
Message
NMT
Battery
Depleted
Skin resistance too
high
NMT
Calibration
Failed
NMT Internal Err.
NMT Comm. Fail
NMT Surface/Needle
Cable Not Mounted
NMT
Acc.
Sensor
Cable Not Mounted
NMT Temp. Cable
Not Mounted
NMT
Acc.
Sensor
Signal Out Of Range
NMT Battery Low
Cause
External device battery is
depleted.
Because electrode position is
wrong, electrode is not well
fixed on skin or patient skin
resistance
resistance is too high, and
then measurement can't be
taken.
Electrode position is wrong,
frequency and sensitivity is
not proper for patient type.
External device's internal
hardware or system error
occurs.
V-Link
communication failure.
Surface cable and needle
cable are both not mounted.
External device's acc cable
is not mounted.
Temp Cable is not mounted.
External device's acc sensor
signal is too high to exceed
rated measurement range.
External device battery is
low.
Alarm Level
Medium
problem,
Medium
Medium
Medium
module
High
Low
Low
Low
Low
Medium
- 69 -
Alarm Information
Action Taken
Replace or charge
the battery.
Check the electrode
and cable to confirm
connection
to
patient, the monitor
and external device
cable
are
well
connected.
Check the electrode
and patient setup to
make sure electrode
is well connected
and patient setup is
correct.
Check
external
device.
Stop
measuring
function of NMT
module, and notify
biomedical engineer
or
manufacturer's
service staff.
Connect
surface
cable
and
needle
cable
Connect
sensor
cable.
Connect
TEMP
cable.
Make
sure
acc
sensor
electrode
connection is well,
and reduce patient
finger motion.
Replace or charge
the battery.