EDAN elite V5 User Manual page 71

Patient monitor
Table of Contents

Advertisement

Patient Monitor User Manual
Message
RESP
Cardiac
Artifact
SpO
2
SpO
Sensor Off
2
SpO
Comm Fail
2
SpO
No Sensor
2
SpO
Sensor Err
2
Cause
No RESP waveform can be
detected due to apnea or
shallow breathing of the
patient.
SpO
sensor
may
2
disconnected
from
patient or the monitor.
SpO
module failure or
2
communication failure
No
SpO
sensor
2
connected to the monitor.
Malfunction in the SpO
sensor or in the extension
cable.
Alarm Level
High
be
the
Low
High
was
Low
2
Low
- 56 -
Alarm Information
Action Taken
Check whether the
patient is breathing
normally.
Take
measures to help the
patient
breathe
normally
when
necessary.
If
the
patient is breathing
normally,
try
to
adjust the electrode
position
on
the
patient in order to
reduce
the
interference
of
cardiogenic artifact.
Make
sure
the
sensor
is
well
connected
to
the
patient's finger or
other parts. Make
sure
the
monitor
and cables are well
connected.
Stop
using
measuring function
of SpO
module,
2
and
notify
biomedical engineer
or
manufacturer's
service staff.
Make
sure
the
monitor and sensor
are well connected,
reconnect
the
sensor.
Replace the SpO
2
sensor
or
the
extension cable.