Contents displayed in the window:
1
Make sure that the computational constant for the measurement is appropriate to the
catheter used.
2
Before a C.O. measurement is initiated, check the accuracy of patient setup. The
calculation of C.O. is related to the patient height, weight, and catheter computation
coefficient; therefore, incorrect input will lead to error in calculation.
NOTE:
1
The blood temperature alarm will not function during C.O. measurement. It will
resume automatically when the measurement is over.
2
It is strongly recommended that the user must push the injector within four seconds
after pressing the Start button.
3
It is strongly recommended that you wait at least 1 minute (or longer depending on
the patient's clinical condition) before starting the next measurement.
Window for C.O. Edit
Six curves of the six measurements and C.O. value
Average value of C.O.
Average value of CI
WARNING
- 146 -
Monitoring C.O.