EDAN elite V5 User Manual page 74

Patient monitor
Table of Contents

Advertisement

Patient Monitor User Manual
Message
NIBP Interference
NIBP
Leak
Test
Error
TEMP
TEMP T1 Sensor Off
TEMP T2 Sensor Off
Excessive T1
Excessive T2
TEMP Comm Fail
T1
Calibration
Failed
T2
Calibration
Failed
IBP
YY Sensor Off (YY
stands for the IBP
label name: Art, PA,
CVP, RAP, LAP, ICP,
P1 and P2)
Cause
Signal noise is too large or
pulse rate is not regular due
to the patient movement.
Fail to deflate normally
during the leak test, so
NIBP leak test cannot be
finished.
Temperature cable of TEMP
channel1
may
disconnected
from
monitor.
Temperature cable of TEMP
channel2
may
disconnected
from
monitor.
TEMP1 measuring value is
beyond measuring range.
TEMP2 measuring value is
beyond measuring range.
TEMP module failure or
communication failure.
T1 calibration failed.
T2 calibration failed
IBP sensor falls off.
Alarm Level
Low
Low
be
Low
the
be
Low
the
High
High
High
High
High
Medium
- 59 -
Alarm Information
Action Taken
Make sure that the
patient
under
monitoring
is
motionless.
Test again. If the
problem still exists,
contact your service
personnel.
Make sure that the
cable
is
properly
connected
Make sure that the
cable
is
properly
connected.
Check
sensor
connection
and
patient condition
Stop
measuring
function of TEMP
module, and notify
biomedical engineer
or
Manufacturer's
service staff.
Please
check
whether the module
works properly.
Check
the
sensor
connection
and
reconnect the sensor.