EDAN elite V5 User Manual page 263

Patient monitor
Table of Contents

Advertisement

Patient Monitor User Manual
☆ Electrode
Offset
Potential Tolerance
Auxiliary
Current
(Leads off detection)
☆ Recovery
Time
After Defibrillation
Leakage Current of
Patient
Scale Signal
☆ System Noise
Multichannel
Crosstalk
☆ Frequency
Impulse Response
Sampling Frequency
Sampling
Channel
Switch Time
A/D Precision
Electrosurgical
Interference
Suppression
☆ ESU Protection
Minimum Input Slew
Rate (Lead II)
☆ Baseline
Reset
Time
± 500 mV
Active electrode: < 100 nA
Reference electrode: < 900 nA
< 5 s (measured without electrodes as IEC60601-2-27:2011, Sect.
201.8.5.5.1 requires.)
< 10 μA
1 mV PP, accuracy is ±5%
< 30 μVPP (RTI)
≤ 5% of the input signal
Complied with IEC 60601-2-27: 2011, Sect. 201.12.1.101.5.
Frequency response:
and
Input a 5 Hz, 1 mV sine wave signal, and the output signal amplitude
remains within the range of 71% to 110% at 0.67 Hz and 40 Hz.
Input a 1 Hz, 1.5 mV 200 ms triangular wave input signal, and the
output shall be within 11.25 mm~15 mm.
Impulse response:
Displacement value: ≤ 0.1 mV
Slope: ≤ 0.3 mV/s following the end of the pulse.
Complied with IEC 60601-2-27: 2011, Sect. 201.12.1.101.8.
1000 Hz
< 80 μS
24 Bits
Test according to ANSI/AAMI EC13:2002, Sect. 5.2.9.14. Complied
with ANSI/AAMI EC13:2002, Sect. 4.2.9.14.
Cut mode: 300 W
Coagulation mode: 100 W
Restore time: ≤ 10 s
> 2.5 V/s
< 3 s
- 248 -
Product Specifications