EDAN elite V5 User Manual page 85

Patient monitor
Table of Contents

Advertisement

Patient Monitor User Manual
Message
NMT Temp. Value
Out Of Range
PTC error
Other external devices
XX Comm. Failed
(XX represents one of
the external devices
connected.)
V-Link Abnormity
Others
Battery Low
Battery1 Error
Battery2 Error
Recorder
Out
Paper
Recorder
Probe
Overheated
Cause
External
device's
sensor signal over flow and
exceed rated measurement
range.
In PTC mode, the patient
response
in
stimulation.
External module failure or
communication failure.
Communication
between
monitor
V-Link.
Battery Low
Malfunction in Battery 1
Malfunction in Battery 2
Of
Recorder Out Of Paper
The probe of recorder is
overheated.
Alarm Level
temp
Low
the
first
Low
High
failure
Medium
and
High
Low
Low
Low
Low
- 70 -
Alarm Information
Action Taken
Make
sure
temp
sensor
is
well
connected
and
temperature
is
within measurement
range,
Check the patient
response and drug
dose is proper.
Unplug the V-Link
module, insert the
serial port cable, and
then reinsert V-Link
module.
If
the
problem still exists,
please
notify
biomedical engineer
or
manufacturer's
service staff.
Please contact the
manufacturer
and
replace
V-Link
module.
Change the batteries
or
charge
the
batteries.
Replace the battery
and
restart
the
monitor.
If
the
problem
persists,
notify
the
manufacturer's
service staff.
Please install the
paper
Stop recording and
retry after the probe
cools.