EDAN elite V5 User Manual page 171

Patient monitor
Table of Contents

Advertisement

Patient Monitor User Manual
3 mmHg. If it has decreased more there is a major leakage in the ISA unit or in the Nomoline.
Do not operate the ISA if there is a major leakage in the unit.
17.3.1.3 System Setup for Analyzer
If your system is using the plug-in and measure ISA analyzer, please follow the setup instructions
below:
1. Connect the ISA analyzer interface cable to the monitor.
2. Connect a Nomoline sampling line to the ISA analyzer input connector.
3. Connect the gas sample exhaust port to a scavenging system or return the gas to the patient
circuit.
4. Power up the monitor.
5. A green LED indicates that the ISA analyzer is ready for use.
6. Perform a pre-use check as described in section Perform a pre-use Check.
17.3.1.4 Zeroing
The infrared module needs to establish a zero reference level for the CO
agent gas measurement. This zero calibration is here referred to as "zeroing".
ISA analyzer performs zeroing automatically by switching the gas sampling from the respiratory
circuit to ambient air. The automatic zeroing is performed every 24 hours, and takes less than 10
seconds for ISA analyzer.
If the ISA analyzer is fitted with an oxygen sensor, the automatic zeroing will also include room
air calibration of the oxygen sensor.
Since a successful zeroing requires the presence of ambient air (21% O
ensure that the ISA analyzer is placed in a well ventilated place. Avoid breathing near the
ISA analyzer before or during the zeroing procedure.
17.3.1.5 Cleaning
The ISA sidestream gas analyzers and Nomoline Adapter can be cleaned using a cloth moistened
(not wet) with max 70% ethanol or isopropyl alcohol.
To prevent cleaning liquids and dust from entering the ISA gas analyzer through its LEGI
connector, keep the Nomoline Family sampling line connected while cleaning the analyzer.
Never immerse the ISA sidestream gas analyzer in liquid.
WARNING
CAUTION
- 156 -
Monitoring AG
, N
O and anesthetic
2
2
and 0% CO
2
2
),

Hide quick links: