EDAN elite V5 User Manual page 266

Patient monitor
Table of Contents

Advertisement

Patient Monitor User Manual
Ventricular
Tachycardia
1 mV 206 bpm
Ventricular
Tachycardia
2 mV 195 bpm
Response
Time
Heart Rate Meter to
Change in HR
Tall
T-wave
Rejection
Accuracy
of
Heart
Rate
Meter
Response to Irregular
Rhythm
Time to Alarm for
Heart
Rate
alarm
conditions
Arrhythmia analyses
Gain 0.5: 10 s
Gain 1.0: 10 s
Gain 2.0: 10 s
Gain 0.5: 10 s
Gain 1.0: 10 s
Gain 2.0: 10 s
of
HR range: 80 bpm to 120 bpm
Range : Within 11 s
HR range: 80 bpm to 40 bpm
Range : Within 11 s
Complied with IEC 60601-2-27: 2011, Sect. 201.12.1.101.17
minimum recommended 1.2 mV T-Wave amplitude
Complied with IEC 60601-2-27: 2011, Sect. 201.7.9.2.9.101 b) 4), the
and
HR value after 20 seconds of stabilization is displayed as follows:
Ventricular bigeminy: 80 bpm ±1 bpm
Slow alternating ventricular bigeminy: 60 bpm ±1 bpm
Rapid alternating ventricular bigeminy: 120 bpm ±1 bpm
Bidirectional systoles: 91 bpm ±1 bpm
Asystole alarm: ≤ 10 s
HR low alarm: ≤ 10 s
HR high alarm: ≤ 10 s
Asystole
Vent Rhythm
Tachy
Irr Rhythm
Pacer not Pacing
VEB
IPVC
Pauses/min High
PAC Bigeminy
ExtremeBrady
Sustain VT
V-Fib/V-Tach
PVC Bigeminy
R on T
Brady
Vent Brady
Run PVCs
Non-Sustain VT
Pause
PVCs High
PAC Trigeminy
ExtremeTachy
- 251 -
Product Specifications
Couplet
PVC Trigeminy
PVC
Missed Beat
Pacer not Capture
Acc. Vent Rhythm
Multiform PVCs
Afib
Low Voltage(Limb)
Wide QRS Tachy
V-Tach