EDAN elite V5 User Manual page 77

Patient monitor
Table of Contents

Advertisement

Patient Monitor User Manual
Message
AG
AA
Unreliable
AG Replace Adapter
AG TEMP Out Of
Range
Sample
Line
Occluded
AG No Adapter
No Sample Line
AG Occlusion
Check
Watertrap/Sample
Line
AG
Change
Watertrap
Watertrap will be
full
Cause
1) Mainstream: The airway
adapter
was
without a zeroing.
2) More than 2 anesthetic
agents are present in the
Id
breathing circuit.
3) High concentrations of
solvents, cleaning agents or
other interfering gases are
present in the breathing
circuit.
Replacement of the adapter
is required.
The temperature of the AG
module
exceeds
specified
temperature range.
The
sampling
occluded.
No adapter is connected.
No
sampling
connected.
AG module sample line
occluded
Watertrap or sample line
falls off.
Malfunction in watertrap
Watertrap will be full.
Alarm Level
replaced
Medium
Medium
the
High
working
tube
is
Medium
Medium
tube
is
Medium
High
Low
Medium
Medium
- 62 -
Alarm Information
Action Taken
1)
Perform
a
zeroing
after
replacing
the
adapter.
2)
Reduce
the
number
of
anesthetic
agent
types.
3)
Replace
the
sampling tube or
reduce
the
interfering gases.
Replace the adapter.
Make sure the AG
module
is
used
within the specified
temperature range.
Replace
the
sampling tube.
Connect the adapter
correctly.
Connect
the
sampling
tube
correctly.
Replace
the
sampling line.
1) Check whether
water trap is
installed normally.
2) Check whether
sample line is
installed normally.
Replace
the
watertrap.
Replace
the
watertrap.