EDAN elite V5 User Manual page 289

Patient monitor
Table of Contents

Advertisement

Patient Monitor User Manual
Purging
Flow
Range
☆ Accuracy
Resolution
☆ Tidal Volume
☆ Range
☆ Accuracy
Resolution
☆ Airway Pressure
☆ Range
☆ Accuracy
Resolution
☆ AwRR
☆ Range
☆ Accuracy
Automatic. Occurs during exhalation.
Adult and Pediatric: 2.5 second duration per line at 10 minute intervals
Neonatal: 1.5 second duration at 3 minute intervals
Adult
Pediatric
Neonatal
Adult
Pediatric
Neonatal
1.0 L/min
Adult
Pediatric
Neonatal
Adult
Pediatric
Neonatal
Adult/Pediatric
Neonatal
Adult/Pediatric/Neonatal
Adult/Pediatric/Neonatal
Adult/Pediatric/Neonatal
Adult/Pediatric/Neonatal
Adult/Pediatric/Neonatal
2.0 L/min to 180 L/min
0.75 L/min to 100 L/min
0.25 L/min to 30 L/min
0.5 L/min or ± 3% of reading,
whichever is greater
0.25 L/min or ± 3% of reading,
whichever is greater
0.125 L/min or ± 3% of reading,
whichever is greater
40 mL to 2500 mL
6 mL to 750 mL
2 mL to 100 mL
± 10.0 mL or ± 5% of reading,
whichever is greater
± 3.0 mL or ± 5% of reading,
whichever is greater
± 1.0 mL or ± 5% of reading,
whichever is greater
1.0 mL
1.0 mL
-120 cmH
0.5 cmH
whichever is greater
1 cmH
O
2
2 rpm to 150 rpm
± 1 rpm
- 274 -
Product Specifications
O to 120 cmH
O
2
2
O or ± 2% of reading,
2