Patient Monitor User Manual
Message
AG Agent Mixture
CO
2
CO
Comm Fail
2
CO
Sensor
Over
2
Temp
CO
Sensor Faulty
2
CO
Zero Required
2
CO
Check Adapter
2
CO₂ Out Of Range
Cause
Mixture agents are detected,
but the monitor cannot
calculate MAC because of
low concentration.
CO
module
failure
2
communication failure
CO
sensor
temperature
2
exceeds +40 °C.
CO
module failure
2
Zero calibration failure
1) For the Respironics CO
module: The cannula is off
or disconnected.
2) For the EDAN CO
module: The water trap is
disconnected or not properly
connected.
The
CO
concentration
2
exceeds the accuracy range
of CO
module.
2
Alarm Level
Medium
or
High
High
High
Low
2
Low
2
High
- 63 -
Alarm Information
Action Taken
Check agents'
concentration ratio.
Check if the water
tray has been fixed.
Stop
using
measuring function
of
CO
module,
2
notify
biomedical
engineer.
Disconnect
the
sampling cannula or
adapter
from
the
airway; initiate the
zeroing
before
making sure that no
expired air is inside
the
sampling
cannula and adapter.
1)
For
the
Respironics
CO
2
module:
Check
whether the adapter
is
properly
connected
or
replace the adapter.
2) For the EDAN
CO
module:
2
Properly
connect
the water trap.
Reduce
CO
2
concentration.