10 Operation
155
Fig. 10.11: Gas component
Unit Conversion
An implemented formulary offers additional options for a conversion. Values from
the PGC can be converted according to the requirements. For example to multiply
the methane concentration (0.94) with 100 (94%) in coordinate AX-46 has to be
selected F8252 * 100.
Fig. 10.12: Applying the unit conversion
Rules for distributing the gas components
Not all possible gas components can be filled. Then these gas components are to
be distributed according to the distribution rules.
"Treatment of the gas data" on page 32
It might be possible that there is no input field in the USM-GT-400 for a measured
gas component from the PGC, for example, Neo-pentane (register 8264). Neo -
pentane will then be added to N-pentane (register 8268). Coordinate AZ-15 then
has to be formulated as F8264 + F8268.
Fig. 10.13: Applied distribution rule
Manual USM GT400 · EN10 · May 2021