10 Operation
148
▪ Conclude work
1
Switch off the system power supply.
2
Open the cover of the ultrasonic electronics.
"Opening the cover of the ultrasonic electronics" on page 145
3
Press calibration switch (A) downwards to close it.
4
Close the cover of the ultrasonic electronics.
"Screw down the cover of the ultrasonic electronics" on page 146
Manual USM GT400 · EN10 · May 2021
If the lock is not shown as open, check the switch position of
the calibration switch.
3
Activate the button with the magnet.
The value of the parameter can be changed.
Figure 10-1 on page 139
4
Activate the button with the magnet in order to position the cur-
sor on the value of the parameters to be changed.
5
Activate the button with the magnet in order to set the desired
value.
6
Activate the button with the magnet in order to confirm the
desired value.
When the parameter is changed, you have to move the
calibration switch back to the closed position.
1
Calibration switch
Fig. 10.7: Close the calibration switch