RMG USM GT400 Operating Manual page 137

Ultrasonic flowmeter
Hide thumbs Also See for USM GT400:
Table of Contents

Advertisement

▪ Establish connection with the clamping screw connection
1
Union nut with clamping screw connection Ø 6 mm
2
Blind plug (plugged)
Fig. 8.39: Connection options of the pressure connection with clamping screw
1
Unscrew the union nut of the clamping screw connection (1).
2
Remove the blind plug (2).
3
Push the union nut and clamping rings onto the pipe.
4
Push the pipe into the clamping screw connection until the stop.
5
Tighten the union nut in order to fix and seal the pipe.
Usually, Swagelok screw connections (or similar) are used here.
6
Tighten the Swagelok screw connection hand tight.
7
Tighten the Swagelok screw connection using a spanner (wide across flats
14) by a further ¼ turn.
connection
Manual USM GT400 · EN10 · May 2021